F720
LEAK TEST INSTRUMENT
FOR INDUSTRIAL QUALITY CONTROL
ATEQ introduces the F720, part of the brand-new Series 7, a next-generation product line set to revolutionize measurement in factories and laboratories. Its modern design and dynamic interface are engineered to captivate users and elevate testing efficiency. Guided by customer needs, our teams continuously innovate and enhance expertise to provide high-quality solutions and service worldwide.
Dynamic, user-centric touchscreen interface
Secure cloud connectivity for real-time data and documentation access
High-precision, high-repeatability measurements
Worldwide presence and International Expertise
F720 : MEASUREMENT RANGES
| PRESSURE DROP MEASUREMENT | |||
|---|---|---|---|
| Range (Pa) | Standard accuracy* | Optional accuracy (Laboratory calibration on request) | Max. resolution (Pa) |
| 0 – 20 | +/- (2% Rdg + 0.3 Pa) | 0 to 2 Pa: +/- 0.1 Pa Over 2 Pa: +/- (2% Rdg + 0.1 Pa) | 0.001 |
| 0 – 50 | Calibration only in laboratory | +/- (1% Rdg + 0.5 Pa) | 0.01 |
| 0 – 500 | +/- (1% Rdg + 1 Pa) | +/- (0.5% Rdg + 0.5 Pa) | 0.1 |
| 0 – 5000 | +/- (1% Rdg + 10 Pa) | +/- (0.5% Rdg + 5 Pa) | 1 |
*Accuracy: Deviation + Repeatability + Hysteresis.
Rdg: Reading.
F720 : FEATURES
MAIN FEATURES
- Up to 128 programs with 3 different modes: Program, Group (Program sequencing) and Batch (Statistics analysis)
- Fully integrated electronics and pneumatics
- Integrated screenshot capture and export capabilities
- Numerical and graphical data visualization interface
- Flexible parts testing capability
- Rapid filling with a wide test range, from fractions of a cc to several hundred Liters
- Local result history with default archiving
- Multilingual user interface
TEST MODES
- Differential leak test
- Pressure decay leak test
- Sealed part leak test
- Volume test
- Blocage test
FLEXIBLE COMMUNICATION FORMAT
- Ethernet: Modbus TCP/IP, Results server
- Master USB: Parameters and results transfer
- Slave USB: PC supervision
- RS232: printer, modbus RTU
- Option Fieldbus: Profinet, Ethernet/IP
OPTIONS & ACCESSORIES
- 9 inputs / 5 local outputs
- 6 inputs/ 6 outputs 24 V programmable for external automation
- Electronic regulator
- Dual pressure
- 2 pneumatic outputs for control of sealing connectors
- Y valve
- Leak calibrator (CDF7000)
- Master leak
- Pneumatic connectors
- PLC Tool Box
- Barcode reader
TECHNICAL SPECIFICATIONS :
| SPECIFICATIONS | |
|---|---|
| Physical | Weight: 7 kg (15.43 lbs) (Height x Depth x Width) Case dimensions: 144 x 272 x 257 mm Overall dimensions: 156 x 357x 257 mm |
| Interface | User friendly interface: 7” touchscreen and intuitive menu-driven operator controls |
| External electrical supply | 24VDC/2.5A (power adapter provided) |
| Air Supply | 6 bar (87 PSI) ± 15% 1 to 2 bar above test pressure Clean and dry air Air quality standard to be applied (ISO 8573-1) |
| Temperature | Operating: +5°C to +45°C Storage: 0°C to +60°C |
| Environment | ROHS standard |



F720 – F750 – F770 COMPARISON
F720 |
F750 |
F770 |
|||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Product profile | Versatile, next-generation Series 7 | Compact, ideal for limited space | Large format for rack/machine integration | ||||||||||||||||||||||||||||||||||||||||
| Weight | 7 kg | 7 kg | 8 kg | ||||||||||||||||||||||||||||||||||||||||
| Dimensions (H × D × W) | 144 × 272 × 257 mm | 245 × 286 × 174 mm | 134 × 273 × 442 mm | ||||||||||||||||||||||||||||||||||||||||
Other specifications ▼
|
|||||||||||||||||||||||||||||||||||||||||||
F720 : NEW TOUCHSCREEN INTERFACE
F720: FACTSHEETS
Detailed information about the F720 including features, specifications, etc.
ATEQ supports machine builders throughout the machine lifecycle, ensuring performance and long-term success.